Technical Speci cations
Measurements procedures:
Endpoint, Fixed Time, Kinetic, Bichromatic
Blank: Water, Reagent blank or Sample blank
Calibration: Factor, Calibrator and Multipoint
Up to three di erent reagent additions possible for
each method
Samples can be loaded by single test or pro le
• 9 di erent calculation algorithms programmed for lin-
ear and non-linear multipoint analysis
Operation:
Random access 130 tests/hr
Runs clinical chemistry and turbidimetric tests
Large choice of Cypress Diagnostics reagents available
Up to 300 clinical applications programmable
Photometer particulars:
Multi wavelength double beam photometer with nar-
row band interference  lters
Filters (8): 340, 380, 405, 492, 510, 546, 576, 620 nm
Halogen lamp 6V/10W (2000 hours)
Linear range: 0,0005 – 3,5000 ABS
Resolution: 0,0005 ABS
Drift: < 0,0001 ABS
Detector: Silicon Diode
Automatic autozeroing on the sample and reference
channel (double beam)
Option:
Barcode reader for reagents and samples
Sampling:
High precision diluter module
Sample aspiration: 2-99 µl with 1 µl increments
Reagent 1 volume: 3-500 µl with 1 µl increments
Reagent 2 & 3 volumes: 3-330 µl with 1 µl increments
Minimum reaction volume: 220 µl
Maximum reaction volume: 550 µl
Precision: CV < 1% at 2 µl
Liquid level sensor
Preheated sampling line at 37°C
Automatic probe wash cycle in between tests to avoid
cross-contamination
Cuvettes:
120 semi-disposable reading cuvettes (6 racks with 20
positions) with a 6 mm light path
Automatic washing and drying cycle by the washing
station for reusing the cuvettes
Automatic QC of cuvettes after washing
Temperature controlled at 37°C
Capacity:
32 sample positions (4 racks with 8 positions) of 3 ml
sample cups or primary tubes
24 reagent containers (35 ml or 6 ml)
12 positions for controls and calibrators
Operating environment:
In covered and dry places
Temperature: 18°C – 32°C
Relative humidity: 20 %-85 %
Stable surface (no vibrations)
> 10 cm around the instrument to permit air circulation
Weight & dimensions:
53 x 57 x 75 cm (H x W x L)
• 37 Kg
Power:
110/240 V – 47/63 Hz
• 300 W
• Advised:
Electronic Stabilizer (> 1kW Power potential)
No-Break Module (UPS - Uninterrupted Power Supply)
Software features:
Unlimited work list
Historical patient results database
Methods, calibrators and controls database
Method programming includes control parameters
• STAT-samples
Quality Control and visualization and printout of graphs
(Levey-Jennings)
Sample pre-dilution and automatic dilution in rerun
mode
Automatic rerun programmable
Reference values are user de nable for serum and urine
for child, female and male
Sample entry for serum/urine
Multipoint calibration curve (automatic dilution of mas-
ter standard) and relative graphs
Bidirectional Interface with Host Computer
Auto-Diagnostics Program for a system built on di er-
ent modules
Communication: USB 1.1/2.0 Full Speed & RS-232
Operative System: Windows XP, Vista & Windows 7
Language: English, French and other languages upon
request
Screen Resolution: 800 x 600 Pixels
Order code:
• CY008 CYANExpert 130